Error 253: Calibration cartridge programming error
Description
The calibration cartridge inserted on the MPI® DESS® post had not been properly programmed.
Possible causes
| 1. | The cartridge contacts are rusted or dirty. |
| 2. | The cartridge was already programmed or is defective. |
| 3. | The MPI is defective. |
Suggested actions
| 1. | Make sure that the cartridge is completely inserted on the MPI DESS post. |
| 2. | Clean the cartridge contacts. |
| 3. | Maintain a light pressure with your hand on the cartridge. |
| 4. | Try programming the cartridge again. |
| 5. | Try another cartridge. |